Comparison of sixteen serological SARS-CoV-2 immunoassays in sixteen clinical laboratories
38 auth. L. Harritshøj, M. Gybel-Brask, S. Afzal, Pia R. Kamstrup, C. Jørgensen, M. Thomsen, L. Hilsted, L. Friis-Hansen, P. Szecsi, Lise Pedersen, L. Nielsen, C. Hansen, P. Garred, T. Korsholm, S. Mikkelsen, ... K. O. Nielsen, B. Møller, Anne T. Hansen, K. Iversen, P. B. Nielsen, R. Hasselbalch, K. Fogh, J. Norsk, J. Kristensen, K. Schønning, N. Kirkby, A. C. Nielsen, L. H. Landsy, M. Loftager, D. Holm, A. Nilsson, S. Sækmose, Birgitte Grum-Schwensen, B. Aagaard, T. G. Jensen, D. M. Nielsen, H. Ullum, R. Dessau
Serological SARS-CoV-2 assays are needed to support clinical diagnosis and epidemiological investigations. Recently, assays for the large-volume detection of total antibodies (Ab) and immunoglobulin (Ig) G and M against SARS-CoV-2 antigens have been…
Serological SARS-CoV-2 assays are needed to support clinical diagnosis and epidemiological investigations. Recently, assays for the large-volume detection of total antibodies (Ab) and immunoglobulin (Ig) G and M against SARS-CoV-2 antigens have been developed, but there are limited data on the diagnostic accuracy of these assays. This study was organized as a Danish national collaboration and included fifteen commercial and one in-house anti-SARS-CoV-2 assays in sixteen laboratories. Sensitivity was evaluated using 150 serum samples from individuals diagnosed with asymptomatic, mild or moderate nonhospitalized (n=129) or hospitalized (n=31) COVID-19, confirmed by nucleic acid amplification tests, collected 13-73 days from symptom onset. Specificity and cross-reactivity were evaluated in samples collected prior to the SARS-CoV-2 epidemic from > 586 blood donors and patients with autoimmune diseases or CMV or EBV infections. Predefined specificity criteria of [≥]99% were met by all total-Ab and IgG assays except one (Diasorin/LiaisonXL-IgG 97.2%). The sensitivities in descending order were: Wantai/ELISA total-Ab (96.7%), CUH/NOVO in-house ELISA total-Ab (96.0%), Ortho/Vitros total-Ab (95.3%), YHLO/iFlash-IgG (94.0%), Ortho/Vitros-IgG (93.3%), Siemens/Atellica total-Ab (93.2%), Roche-Elecsys total-Ab (92.7%), Abbott-Architect-IgG (90.0%), Abbott/Alinity-IgG (median 88.0%), Diasorin/LiaisonXL-IgG (84.6%), Siemens/Vista total-Ab (81.0%), Euroimmun/ELISA-IgG (78.0%), and Snibe/Maglumi-IgG (median 78.0%). The IgM results were variable, but one assay (Wantai/ELISA-IgM) had both high sensitivity (82.7%) and specificity (99%). The rate of seropositivity increased with time from symptom onset and symptom severity. In conclusion, predefined sensitivity and specificity acceptance criteria of 90%/99%, respectively, for diagnostic use were met in five of six total-Ab and three of seven IgG assays.
Published in medRxiv
19
5 2020